GT6800-12 multi-parameter monitor
1. The monitor adopts the world module. High precision. The parameters are accurate. Reliable quality, stable performance
2 , high brightness and high resolution 12.1 inch color TFT liquid display. The data waveform is clear and easy to see
3. Rotating shuttle operation mode. Chinese and English dual-purpose operation interface. Easy to learn. Easy to use, easy to operate
4 , a seven-channel synchronous full-lead ECG display
5 , five minutes ECG waveform playback, ST segment automatic detection
6 , all parameters 38 is called, when the trend data memory and dynamic waveform capture
7 , can be set Alarm upper and lower limits, automatic sound and light dual alarm
8 , suitable for adults, children, newborns
9 , anti-defibrillation, anti-interference of high-frequency electric knife
10 , remote wired central monitoring network interface
11 , optional plug-in lithium battery. AC and DC dual purpose
12 , optional built-in dual-channel thermal array recorder, real-time printing of waveforms and graphs
13 , widely used in internal medicine, surgery, operating room, obstetrics and gynecology, pediatrics, emergency centers, ICU , CCU
GT6800-10Multi-parameter monitor technical parameters
1, the ECG ECG
lead selection: Full lead, Ⅰ, Ⅱ, Ⅲ, aVF , aVR , aVL , V1--6 chest leads show Gain Select: × . 1 , × 2 , × . 4 , × 0.5 four-speed frequency: diagnostic mode 0.05 ~ 130Hz monitoring mode 0.5 ~ 40Hz operation mode 1 ~ 20Hz common mode signal suppression: diagnostic mode > 90 dB monitoring mode > 100 dB surgery mode > 100 dB scanning speed: 12.5 mm / S , 25 mm / S , 50 mm / S
Three standard files.
Heart rate range: adult 15 ~ 300bpm ( beat / min ) newborn / child 15 ~ 350 bpm ( beat / min ) Heart rate: ± 1% or ± 1bpm , whichever is greater Calibration signal: 1mV (peak-to-peak value, accuracy ± 5% ) ST segment detection: measurement range ( -2.0mV ~ +2.0mV ) baseline recovery time: <3 seconds after defibrillation Alarm mode: sound and light alarm heart rate alarm preset range: (exceeds the preset value alarm) upper limit 40 ~Adjustable between 255 times / minute
The lower limit is adjustable between 20 ~160 times / min.
2. RESP respiration measurement range: adult 7 ~120BrPM ( times / min ), children and newborns 7 ~150 BrPM( times / min ) 2 BrPM( times / min ) ± accuracy : Resolution: 1 BrPM ( times / min ) Gain selection: × 0.5 , × 1 , × 2 , × 4 Four alarm modes: sound and light alarm Alarm preset range: (Alarm exceeding the preset value) upper limit
The
lower limit can be adjusted between 10 ~100 times / minute, and the lower limit can be adjusted between 0 ~99 times / minute.
3. NIBP blood pressure technology: oscillometric method (pulse wave oscillation method) Working mode: manual / automatic measurement unit: mmHg / Kpa overpressure Protection: Double safety protection Adult mode 300 mmHg Child mode 240 mmHg Neonatal mode 150 mmHg Blood pressure measurement range and accuracy: range Adult systolic blood pressure 40 ~ 270mmHg Diastolic blood pressure 10 ~ 215mmHg Mean pressure 20 ~ 235mmHg Child systolic blood pressure 40 ~200mmHg
diastolic blood pressure 10 ~150mmHg
average pressure 20 ~165mmHg
neonatal systolic pressure 40 ~135mmHg diastolic pressure 10 ~100mmHg average pressure 20 ~110mmHg static pressure range: 0 ~300mmHg static pressure accuracy: 3mmHg 5mmHg ; standard deviation 8mmHg ± blood pressure accuracy: average Error alarm mode: audible and visual alarm. Preset range and error: (Alarm when the preset value is exceeded) The upper limit of systolic blood pressure is 40 ~ 250mmHg ; the lower limit is 10 ~ 220mmHg The upper limit of diastolic blood pressure is 20 ~ 250mmHg ; lower limit 10 ~ 220mmHg
4. SPO2 blood oxygen saturation
2 DIGIT 0% ~ 69% : not defined ± range: 70 ~ : 20 % ~ 69% : not defined ± blood oxygen saturation measurement error: 70 ~ : Resolution: 1% Pulse frequency: 20 ~ 300 times / min. Alarm mode: Sound and light alarm. Alarm preset range and error: (Alarm exceeding the preset value) upper limit 20% ~; lower limit 10% ~99% ; 5. TEMP Body temperature technique:
Highly sensitive thermistor probe
body temperature measurement range: 0 ℃ ~ 45 ℃ resolution: 0.1 ℃ C (excluding sensor error) ° Body temperature accuracy: 0.1 Alarm mode: sound and light alarm alarm preset range and error: (exceed the preset value Alarm) upper limit adjustable between 20.1 ℃ ~45 ℃ lower limit adjustable between 20 ℃ ~44.9 ℃ 6. Printer (optional) specifications: dual-channel thermal array recorder Printing method: real-time or event trigger recording waveform, text resolution : Horizontal: 400dpi / Vertical: 800dpi Note: name, gender, date (year/month/day)/time ( 24 hours) 7. Power supply Built-in battery (optional): fully sealed, maintenance-free, large-capacity lithium battery External AC power: single item AC power supply 220V , 50Hz , current 1A
working temperature: 5 ℃ ~40 ℃ working humidity: ≤ 85% relative humidity, non-condensing storage temperature: -20 ℃ ~+55 ℃ storage humidity: ≤ 95% relative humidity, non-condensing Altitude: 86.0kPa ~106.0 kPa ( Working status) Power consumption of the whole machine: not more than 70VA